P/L3 Statistical pulsed signal characterization Procedure

Objectives

Characterize the capacity of the RFI Monitoring system to detect pulsed signals by statistical algorithms.

Steps to follow

  • Turn on equipment

    • Power supply
    • Signal generator
  • Configure to desired parameters

    • Power supply
      • Set Voltage to 3.3V
      • Set current limit to 0.3A
    • Signal generator
      • Set output frequency to some IF Band frequency periodic pulsed signals
  • Set up nucleo board and PC

    • Make sure the proper code is loaded into the nucleo board
  • Make connections

    • Connect the power supply to the payload:
      • Vcc to PoL (RED)
      • GND to GND (BLACK)
    • Connect the DAC output to the DAC_In (GREEN)
    • Connect the ADC input to the ADC_Out
    • Load the IF Input with a 50 Ohm perfectly matched load.

    Captura.PNG

    • Connect PC to ground (plug in the charger).
    • Connect necleo board to computer by USB.
  • Power on power supply.

  • Begin the measurements procedure

    • Run the flashed program in Debug mode.
    • Once the program stops at the breakpoint, collect the calculated mean and plot it in a data processing software like Matlab.
  • Analyze the collected data.