P/L3 Statistical pulsed signal characterization Procedure Objectives Characterize the capacity of the RFI Monitoring system to detect pulsed signals by statistical algorithms. Steps to follow Turn on equipment Power supply Signal generator Configure to desired parameters Power supply Set Voltage to 3.3V Set current limit to 0.3A Signal generator Set output frequency to some IF Band frequency periodic pulsed signals Set up nucleo board and PC Make sure the proper code is loaded into the nucleo board Make connections Connect the power supply to the payload: Vcc to PoL (RED) GND to GND (BLACK) Connect the DAC output to the DAC_In (GREEN) Connect the ADC input to the ADC_Out Load the IF Input with a 50 Ohm perfectly matched load. Connect PC to ground (plug in the charger). Connect necleo board to computer by USB. Power on power supply. Begin the measurements procedure Run the flashed program in Debug mode. Once the program stops at the breakpoint, collect the calculated mean and plot it in a data processing software like Matlab. Analyze the collected data.