Front-end Gain calibration Procedure

Objective

Measure the total system gain at the output of the RSSI and equalize it.

Steps to follow

  • Turn on equipment

    • Spectrum analizer
    • Power supply
  • Configure to desired parameters

    • Spectrum analizer
      • Set center frequency in the middle of VCO output band (6250MHz)
      • Set span to match the VCO output band (1400MHz)
      • Set RBW and VBW to an adecuate level
    • Power supply
      • Set Voltage to 3.3V
      • Set current limit to 0.3A
  • Set up nucleo board and PC

    • Make sure the proper code is loaded into the nucleo board
    • Connect the power supply to the payload:
      • Vcc to PoL (RED)
      • GND to GND (BLACK)
    • Connect the DAC output to the DAC_In (GREEN)
    • Connect the ADC input to the ADC_Out
    • Load the IF Input with a 50 Ohm perfectly matched load.
    • Connect the spectrum analizer to the switched measuring pin (PURPLE)

      Make connections

    Top_K-Band_Testing_Connections.PNG

    • Connect PC to ground (plug in the charger)
    • Connect necleo board to computer by USB
  • Power on power supply

  • Begin the measurements procedure

    • Run the flashed program in Debug mode
    • Once the program stops at the breakpoint, collect the calculated mean and plot it in a data processing software like Matlab.
  • Realize the calibration

    • Calculate the mean of the collected data and substract it from each frequency bin mean. This will be the deviation.
    • Extract the trend line from the deviation.
    • Subtract trend line function from the received RSSI value. This will be the calibrated measurement.
  • Reboot the program and check if the equalization is now correct. If it’s not correct repeat the previous step